The specular echo originates from relatively large, strongly reflective, regularly shaped objects with smooth surfaces. These intense reflections are angle dependent, and are described by reflectivity equation. This type of reflection is called specular reflection (i.e. IVS, valves). See also False Distance Artifact, and Scattered Echo. •
Stress echocardiograms are used for detection of coronary artery disease, or to determine the cardiac performance. Stress echocardiograms are less performed to evaluate pulmonary artery pressures, pulmonary hypertension or the significance of valvular heart disease. Stress increases the degree to which the heart contracts. After a myocardial infarction there will be a region of the heart muscle that contracts abnormal at rest. This area may worsen with stress. A coronary artery blockage most often do not impairs the function of the heart at rest. With stress, a region of the heart does not receive enough blood to work effectively and wall motion abnormalities occur. The echocardiographer compares rest and exercise and can determine the presence and severity of a coronary artery blockage. Stress echocardiograms involve: A bicycle stress echocardiogram involves transthoracic echocardiography performed at the rest baseline and after or during different stages of physical exercise. A dobutamine stress echocardiography uses the drug dobutamine instead physical exercise. Transthoracic echocardiograms are routinely performed during stress and rest. Cardiovascular stress represents a minimal risk to the patient. See also Transesophageal Echocardiography, Echocardiography, M-Mode, Curved Transducer, Doppler Ultrasound, History of Ultrasound and History of Ultrasound Contrast Agents. ![]() Further Reading: News & More: •
Subdicing is a technique used to overcome grating lobes. With subdicing each major transducer element is divided into smaller parts, each one being a half wave length.
•
Ultrasound waves are created at harmonics of the delivered frequency.
Subharmonic imaging uses the harmonic oscillation of a system at a frequency that is a simple fraction of its fundamental frequency. The subharmonic response frequencies has half of the fundamental frequency.
The second subharmonic has a half fundamental frequency of one half the frequency, and so on.
See also Harmonic Imaging and Superharmonic Imaging. •
Submicron ultrasound contrast agents are gas-filled, double-walled microspheres with a diameter smaller than 1 μm that rupture when exposed
to ultrasound energy at megahertz frequencies. These agents differ from traditional ultrasound contrast microbubbles in that the submicron bubbles may serve as extravascular agents. They are small enough to travel through the lymphatic system and to be extravasated from tumor neovasculature. The detection of these agents is limited by their hard shell, which requires high-pressure ultrasound insonation
for shell rupture and excitation of the gas bubble. After shell rupture, the gas diffuses rapidly from submicron sized agents. The optimal processing of each echo is important.
|