Bubble Rupture
Ultrasound at the microbubble resonance frequency can cause bubble rupture at high acoustic power (mechanical index (MI) greater than 0.5). The result is a transient high-amplitude, broadband signal containing all frequencies, not only the harmonics. It will create a strong signal in B-mode or a short-lasting multicolored, mosaic-like effect in color Doppler sonography.
Several terms for this typical signal have been used, e.g. induced or stimulated acoustic emission, loss of correlation imaging and sono-scintigraphy.
Cavitation
Cavitation is any activity of highly compressible transient or stable microbubbles of gas and/or vapour, generated by ultrasonic power in the propagation medium. Cavitation can be described as inertial or non-inertial. Inertial cavitation has the most potential to damage tissue and occurs when a gas-filled cavity grows, during pressure rarefaction of the ultrasound pulse, and contracts, during the compression phase. Collapses of bubbles can generate local high temperatures and pressures. Transient cavitation can cause tissue damage.
The threshold for cavitation is high and does not occur at current levels of diagnostic ultrasound. The introduction of contrast agents leads to the formation of microbubbles that potentially provide gas nuclei for cavitation. The use of contrast agents can lower the threshold at which cavitation occurs.

Types of cavitation:
Acoustic cavitation - sound in liquid can produce bubbles or cavities containing gas or vapour.
Stable cavitation - steady microbubble oscillation due to the passage of a sound wave.
Transient cavitation - short-lived cavitation initiated by the negative pressure of the sound wave.

Color Map
In a color map, colors are allocated to Doppler shift frequencies corresponding to flow. Color (colour, Brit.) maps may also display Doppler amplitude, signal power, variance, or the gray levels of the B-mode image.

See also Color Amplitude Imaging, Color Priority, and Color Saturation.
Color Saturation
Color saturation is a characteristic used for example to describe the vividness of a color, or the gradation of hue from unsaturated (white) to fully saturated (100% of the given color).

See also Directional Indicators, and Power Map.
Contrast Enhanced Doppler Imaging
Contrast agents improve the sensitivity of vascular Doppler ultrasound, for example in cerebrovascular sonography or examinations of deep abdominal vessels. They also enlarge the role of transcranial Doppler. Microbubbles can be used with various modes e.g., color and power Doppler imaging, as well as pulsed-wave Doppler to increase the signal intensity. However, the ultrasound system must be suitable for contrast enhanced technology.
Microbubbles usually stay within the vascular space; nevertheless, the contrast enhancement is limited to 2−6 minutes caused by physiologic clearance and bubble destruction.
Depended on the application, contrast agents can be administered with a different injection rate e.g., bolus injection, slow injection, or continuous infusion. Stable, homogeneous, and prolonged enhancement can be obtained with perfusion, lasting until the infusion is stopped.

See also Cerebrovascular Ultrasonography, Multiple Frame Trigger.
Result Pages :
 
Share This Page
Facebook
Twitter
LinkedIn
Look
      Ups
Medical-Ultrasound-Imaging.com
former US-TIP.com
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology TIP • Medical-Ultrasound-Imaging
Copyright © 2008 - 2024 SoftWays. All rights reserved.
Terms of Use | Privacy Policy | Advertise With Us
 [last update: 2023-11-06 01:42:00]