Nyquist Frequency
The Nyquist frequency describes the highest frequency that can be sampled in a pulsed ultrasound system. To record higher frequencies the zero baseline can be shifted in the desired direction.

See also Nyquist Limit, and Sampling Rate.
Nyquist Limit
According to Shannon's sampling theorem, the sampling frequency should be twice the frequency being sampled. The nyquist frequency is the maximum frequency that can be sampled without aliasing. In ultrasound imaging, it is defined as half of the pulse repetition frequency.
NF = PRF/2 (nyquist frequency = pulse repetition frequency/2)
This is the so-called Nyquist limit. If the velocity of flow exceeds the Nyquist limit, the direction and velocity are inaccurately displayed and appear to change direction. Color flow Doppler capitalizes on this effect. This allows detecting flow disturbances from laminar to turbulent flow.

See also Aliasing Artifact, Repetition Rate, and Sampling Rate.
Continuous Wave Doppler
(CWD) Continuous wave (CW) Doppler is an ultrasound imaging mode, which records blood flow velocities along the length of the beam. Continuous wave Doppler uses different crystals to send and receive the signal. The transducer operating in continuous wave mode utilizes one half of the elements and is continuously sending sound waves of a single frequency while the other half is continuously receiving the reflected signals.
The advantages of a continuous wave transducer are a high sensitivity and no Nyquist limit. CW Doppler does not alias but has no depth precision and large gate. The beat frequency is the Doppler shift. CW Doppler echocardiography employs this technique to record the flow of blood through the cardiovascular system.

See also Cross Talk, Periorbital Doppler, and Mirror Artifact.
Sampling Rate
The sampling rate is the number of samples taken per unit of time. In digital signal processing (Nyquist theory) it is necessary to sample twice as fast as the highest frequency component.
Result Pages :
 
Share This Page
Facebook
Twitter
LinkedIn

Look
      Ups
Medical-Ultrasound-Imaging.com
former US-TIP.com
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology TIP • Medical-Ultrasound-Imaging
Copyright © 2008 - 2024 SoftWays. All rights reserved.
Terms of Use | Privacy Policy | Advertise With Us
 [last update: 2023-11-06 01:42:00]