Searchterm 'Line Density' found in 7 articles
1
term [
] - 2 definitions [
] - 4 booleans [
]
Result Pages :
CTS-485
www.siui.com/english/product/485.htm From SIUI Inc.;
'SIUI's newly developed CTS-485 is a portable ultrasound system with advanced technology. With its powerful features, such as 256 grayscale, cineloop, RS232C interface and broadband high-density, multi-frequency probes, the system is designed for professional applications in cardiology, abdominal, small parts, liver, gallbladder, kidney, obstetrics, gynecology, peripheral vascular, etc. The CTS-485 has passed FDA clearance and CE Marking.'

Device Information and Specification
APPLICATIONS
CONFIGURATION
Portable, gray scale(256)
Linear and convex
PROBES STANDARD
1 * 2.5MHz ~ 5.0MHz multifrequency convex probe
2.5MHz to 10.0MHz broad band, trifrequency
IMAGING OPTIONS
Multi zoom rate and depth shift
OPTIONAL PACKAGE
Different probes such as linear, convex, vaginal and rectal probes; Cart; Image communication function and RS-232C interface
H*W*D m
0.26 * 0.3 * 0.41
WEIGHT
11 kg (main unit)
POWER REQUIREMENT
AC 220V/110V, 50Hz/60Hz
POWER CONSUMPTION
0.1 KVA
Acoustic Lens
The acoustic lens is placed at the time the transducer is manufactured and cannot be changed. The acoustic lens is generally focused in the mid field rather than the near or far fields. The exact focal length varies with transducer frequency, but is generally in the range of 4-6 cm for a 5 MHz curved linear probe and 7-9 cm for a 3.5 MHz curved transducer.
Placing the elevation plane (z-plane) focal zone of the acoustic lens in the very near or far field would improve the beam width at precisely those depths. However, this would degrade the beam width to a much greater and unacceptable degree at all other depths.
There are some chemicals in ultrasound couplants that can degrade the acoustic lens, destroy bonding, or change the acoustic properties of the lens. Problematic chemicals include mineral oil, silicone oil, alcohol, surfactants, and fragrances. Fragrance can affect the transducer's acoustic lens or face material by absorption over time into elastomer and plastic materials, thus changing the material's weight, size, density, and acoustic impedance. Surfactants can degrade the bond between the lens and the piezoelectric elements and contribute to the accelerated degeneration of the lens.

See also Retrolenticular Afterglow.
Result Pages :
 
Share This Page
Facebook
Twitter
LinkedIn
Look
      Ups
Medical-Ultrasound-Imaging.com
former US-TIP.com
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology TIP • Medical-Ultrasound-Imaging
Copyright © 2008 - 2024 SoftWays. All rights reserved.
Terms of Use | Privacy Policy | Advertise With Us
 [last update: 2023-11-06 01:42:00]