Coded Excitation
Increasing the frequency of the transmitted power improves the image quality of ultrasound, but the improvement in resolution results in a decreased signal to noise ratio (SNR). Higher acoustic power levels can prevent the loss in SNR, but among other reasons, ultrasound regulations limit this to avoid heating or cavitation.
Coded excitation increase the signal to noise ratio without the loss of resolution by using coded waveforms. Coded excitation allows transmitting a long wide-band pulse with more acoustic power and high penetration of the sound beam.
Color Doppler Imaging
(CDI) Color Doppler imaging depicts the mean frequency shifts of the Doppler signal. Color [colour, Brit.] Doppler imaging is a method for visualizing direction and velocity of movement, such as of blood flow within the cardiac chambers or blood vessels. The flow direction and velocity information gathered by Doppler ultrasonography is color coded onto a gray scale cross-sectional image. The sensitivity of Doppler ultrasound is increased in conjunction with the use of vascular contrast agents.
Direction and blood flow velocity are coded as colors and shades:
Red - flow coming nearer to the probe.
Blue - flow coming away of the probe.

See also Bi-directional Illumination, Color Map.
Color Power Angio
(CPA) Color power angio is a Doppler measurement, which employs the Doppler effect to assess whether blood is moving towards or away from the probe. By calculating the frequency shift of a particular sample volume, its speed and direction can be determined and visualized. This is particularly useful in cardiovascular studies and essential in many areas such as determining reverse blood flow in the liver vasculature in portal hypertension. The CPA information is displayed as a color image superimposed on the 2D gray scale image.

See also Color Power Doppler.
Color Power Doppler
(CPD) CPD is a type of color Doppler to visualize the presence of detectable blood flow. The flow information is based on the amplitude or strength of echoes received from moving cells and not on frequency shifts. Power Doppler is very sensitive to flowing blood but does not provide velocity or directional information.
CPD is less angle dependent than traditional color Doppler, but more sensitive to motion artifacts. Color power angio (CPA) provides better sensitivity to slow flow states.
The color maps for CPD are represented by a single continuous color (colour, Brit.). Because CPD does not provide directional information, no aliasing artifact occurs.

See also Directional Color Power Doppler.
Contrast Harmonic Imaging
(CHI) Contrast harmonic imaging is an ultrasound technique to improve the measurement of blood perfusion or capillary blood flow. Based on the nonlinear properties of contrast agents, CHI transmits at the fundamental frequency but receives at the second harmonic. Contrast enhanced echo signals contain significant energy components at higher harmonics (bubbles acts as harmonic oscillators), while tissue echoes do not. Caused by that contrast signal can be separated from tissue echoes by the characteristic signal.
In combination with the pulse inversion technique, CHI promises very high contrast agent sensitivity with high spatial resolution.

See also Ultrasound Contrast Agent Safety and Hemoglobin.
Result Pages :
 
Share This Page
Facebook
Twitter
LinkedIn
Look
      Ups
Medical-Ultrasound-Imaging.com
former US-TIP.com
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology TIP • Medical-Ultrasound-Imaging
Copyright © 2008 - 2024 SoftWays. All rights reserved.
Terms of Use | Privacy Policy | Advertise With Us
 [last update: 2023-11-06 01:42:00]