Amplitude
The amplitude is the signal height. The amplitude of the ultrasound echo is proportional to the difference of the acoustical impedance caused by different tissue layers. The amplitude decreases with increasing penetration depth of a homogeneous tissue.

See also Color Amplitude Imaging.
Apogee 3500
www.siui.com/english/product/3500.htm From SIUI Inc.;
'Dedicated to ultrasound industry, Shantou Institute of Ultrasonic Instruments, Inc. (SIUI) has launched Apogee 3500, the Digital Color Doppler Ultrasound Imaging System.
With latest imaging technologies, high-definition image quality and excellent practical functions, the Apogee 3500 offers optimal solutions for clinical ultrasonic examination.'

'The Apogee 3500 is available with many high-density, super broadband and multi-frequency probes, such as convex, micro-convex, linear, vaginal, rectal and phased array probes, which are widely applied for different clinical diagnoses, including abdomen (liver, kidney, gall-bladder, pancreas), gynecology (uterus, ovary), obstetrics (early pregnancy, basic OB, complete OB, multi gestation, fetal echo), cardiology (adult and pediatric cardiology), small parts (thyroid, galactophore, testicles, neonate), peripheral vascular and prostate.'

Device Information and Specification
APPLICATIONS
CONFIGURATION
Normal system, color - gray scale(256)
Linear, convex and phased array
PROBES STANDARD
1 * Super broadband convex probe, 1* super-broadband linear probe, 1* phased array probe
2.0 MHz ~ 12.0 MHz, broad band, tri-frequency
B-mode (B, 2B, 4B), M-mode, B/M-mode, real-time compound imaging, panoramic imaging, trapezoidal imaging (linear probes), spectrum Doppler (PWD and CWD), color Doppler flow imaging (CDFI), color power angio (CPA), tissue harmonic imaging (THI)
IMAGING OPTIONS
Real-time ZOOM, zoom rate and position selectable
OPTIONAL PACKAGE
Linear, convex, micro-convex, vaginal, rectal, biplane probes; CD-RW drive; DICOM 3.0 interface and software; ...
H*W*D m
1.29 * 0.52 * 0.75
WEIGHT
110 kg
POWER REQUIREMENT
AC 220V/110V, 50Hz/60Hz
POWER CONSUMPTION
0.6 KVA
Artifact
An image artifact is any image attribute, which is not present in the original imaged object. An image artifact is sometime the result of an improper operation of the imager, and in other times a consequence of natural processes or properties of the human body.
Artifacts in diagnostic ultrasound are a reflection or an echo, which appears on the display and represents the real anatomical structure not correctly. An artifact can be a false, multiple or misleading information introduced by the imaging system or by interaction of ultrasound with the adjacent tissue.

Artifacts in ultrasound can be classified as to their source like e.g.:
physiologic (motion, different sound velocities, acoustical impedances of tissue);
hardware (dimension of the ultrasound beam and the transducer array);

Image artifacts can occur in each medical ultrasound. Then an interpretation of the image is complicated and can eliminate the structural information of objects looking for.

See also Ultrasound Imaging Procedures.
BY 963
By Byk-Gulden (now Altana);
The transpulmonary echo contrast agent under development, BY 963, promises to be an excellent ultrasound contrast agent for the noninvasive assessment of left ventricular function. Intravenous administration of BY 963 phospholipid microbubbles improves the reproducibility of e.g. quantitative analysis.
Backscattering
Ultrasound waves are reflected when there is a change in acoustic impedance. The larger the change, the more ultrasound is reflected. Microbubbles have an enormous difference in acoustic impedance as compared to surrounding fluid due to the large differences in density, elasticity and compressibility.
At low acoustic power (mechanical index less than 0.1), the mechanism of ultrasound reflection is that of Rayleigh scattering and the microbubbles may be regarded as point scatterers. The scattering strength of a point scatterer is proportional to the sixth power of the particle radius and to the fourth power of the ultrasound frequency;; the echogenicity of such contrast agent is therefore highly dependent upon particle size and transmit frequency. The backscattered intensity of a group of point scatterers is furthermore directly proportional to the total number of scatterers in the insonified volume. The concentration of the contrast medium is of importance.

See also Backscatter Energy, Cross-section Scattering.
Result Pages :
 
Share This Page
Facebook
Twitter
LinkedIn
Look
      Ups
Medical-Ultrasound-Imaging.com
former US-TIP.com
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology TIP • Medical-Ultrasound-Imaging
Copyright © 2008 - 2024 SoftWays. All rights reserved.
Terms of Use | Privacy Policy | Advertise With Us
 [last update: 2023-11-06 01:42:00]