Amplitude Shading
Amplitude shading is a method of reducing the side lobe levels in a transducer array. The shading usually causes the main beam to broaden by applying different voltages to the elements of the array.

See also Analog Output Signal.
Array Transducer
An array transducer is composed of multiple piezoelectric crystal elements arranged in an array. Arrays are transducer assemblies with a row of elements, used to focus the beam.

Types of array transducers:
Linear array transducer = the arrays are arranged along a line.
Curvilinear or curved transducer = the arrays are arranged along a convex curve. A curved array is similar to a linear array except that the image created is a sector-type.
Annular array transducer = the arrays are arranged in concentric circles.
Rectangular array transducer = the arrays are arranged in a rectangular pattern.

See also Amplitude Shading, Transducer Types, and Transducer Assembly.
Side Lobe
Side lobes are secondary and smaller acoustic beams falling outside at predictable angles located around the main lobe.

See also Grating-lobe Artifact, Amplitude Shading, and Apodization.
Result Pages :
 
Share This Page
Facebook
Twitter
LinkedIn

Look
      Ups
Medical-Ultrasound-Imaging.com
former US-TIP.com
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology TIP • Medical-Ultrasound-Imaging
Copyright © 2008 - 2024 SoftWays. All rights reserved.
Terms of Use | Privacy Policy | Advertise With Us
 [last update: 2023-11-06 01:42:00]